Mitutoyo gages with an SPC data port send their readings in Mitutoyo’s Digimatic format. A MicroRidge MobileCollect transmitter or GageWay interface reads that output and passes the reading to your SPC software, a spreadsheet or any program that accepts keyboard input. This page covers every Mitutoyo gage in the MicroRidge Measurement Collection System Builder: which cable each one takes, which models have no output, and how to choose between wireless and wired collection.
Find your Mitutoyo gage in the builderOr describe your application to the System Advisor
Counts come from the MicroRidge Measurement Collection System Builder and update when the gage data does.
How Mitutoyo gages connect
Every Mitutoyo SPC cable has the same 10-pin Digimatic plug on the interface end. Only the gage end changes, and Mitutoyo uses more than a dozen gage-end connector types, each with a letter (A, B, C, D, E, F, G, SA, SF and the angled versions). The connector is set by the gage’s port, so two gages from the same family can need different cables. A coolant-proof caliper and a standard caliper are a common example.
Three things decide the part you need:
- The port on the gage. IP65, IP66 and IP67 gages use sealed connectors (types A, B and G). Standard gages use flat connectors (types C and F). Gages released since 2023 with Mitutoyo’s Digimatic S1 output, such as the ID-C and ID-F indicators with an NX suffix and the QuantuMike micrometers ending in -40, use the new SF and SA connectors, which do not mate with older cables.
- Whether the cable has a data button. Caliper and micrometer cables (types A, B, C, CR and SA) have one, and so do the matching MobileCollect connectors. Indicator and height gage cables do not, so the reading is triggered by a foot or hand switch, the read button on the transmitter, or a command from your software.
- Whether the model has output at all. Mitutoyo sells many gages with and without SPC output, and the difference is in the base part number, not the suffix. For example, the 500-171-30 caliper has output and the 500-196-30 does not. The builder flags every model without output.
Our guide Mitutoyo Digimatic SPC cable types explained covers every connector letter, which cables are physically interchangeable, and the part numbers for 1 m and 2 m cables.
Connections used by Mitutoyo gages
Each row is a gage-side connection, with the MicroRidge parts that read it. The wireless column lists the transmitter and the connector cable for it; the wired column lists the Mitutoyo SPC cable that plugs into a GageWay interface or a Digital Remote.
| Gage connection | Models | Wireless (MobileCollect) | Wired (GageWay) |
|---|---|---|---|
| 05CZA662 Series 293, 350, 468, 395 | 611 |
| |
| 905338 Series 543, 568, 547, 570 | 367 |
| |
| 959149 Series 500, 572, 573, 550 | 343 |
| |
| 937387 Series 568, 468, 293, 227 | 283 |
| |
| 05CZA624 Series 500, 573, 552, 536 | 194 |
| |
| 06AGL011 Series 543, 547 | 98 |
| |
| 936937 Series 518, 178, SJ, 574 | 52 |
| |
| 04AZB512 Series 293, 340, 389 | 29 |
| |
| 936937-EP Series 543 | 22 |
|
|
| RS-232, DB9 female, null modem Series LSM, VL, 518 | 16 |
|
|
| 937243 Series 500, 192 | 15 |
|
|
| Ask MicroRidge Series 518, 568, 983 | 9 | Ask MicroRidge | Not offered |
13 more connections are used by 39 other models: RS-232, DB9 female, 21EAA194, 06AGL111, 937386, RS-232, DB9 male, RS-232, DB25 female, PRO3600, 22806-MIT, RS-232, Digimatic, RS-232, DB9, RS-232, DB25 / Digimatic, RS-232, 10-pin T&B Ansley 609-1027, RS-232, 6-pin modular (RJ-12). The builder has the parts for each.
Prices are U.S. list. The builder adds the base receiver or GageWay interface for your computer and checks each model.
Mitutoyo gage types
Calipers, micrometers and indicators have their own pages with series tables, connector details and every model number:
Height gages, bore gages, depth and thickness gages, Holtest internal micrometers, surface testers and counters are listed below by series. Select a series to open the builder filtered to it.
Micrometers 760
Mitutoyo micrometers: series, cables and every model number
Calipers 627
Mitutoyo calipers: series, cables and every model number
Indicators 340
Mitutoyo indicators: series, cables and every model number
Bore gages 250
| Series | Gage types | Models | Connection |
|---|---|---|---|
| 468 | Holtest, Holtest Set | 116 | 05CZA662, 937387 |
| 568 | Snap-Open Bore Gage, Borematic, Borematic Set | 110 | 937387, 905338 |
| 339 | Tubular Inside Micrometer, Inside Micrometer, Digimatic Tubular Inside Micrometer | 10 | 05CZA662, 937387 |
| 345 | Caliper type inside micrometer | 8 | 05CZA662 |
| 543 | For Bore Gage | 3 | 905338 |
| 511 | Bore Gage | 2 | 905338 |
| 526 | Hole Check | 1 | 905338 |
Height gages 111
| Series | Gage types | Models | Connection |
|---|---|---|---|
| 570 | Height Gage, Height Gage. HDS-40"C | 41 | 905338 |
| 518 | QM Height, Height Gage, Linear Height Gage | 24 | 936937, RS-232, DB9 female, null modem |
| 192 | Height Gage, Height Gage - cable discontinued, Digimatic Height Gage | 22 | 905338, 937243 |
| 515 | Height Gage Master, Digital Height Master | 12 | 937387, 959149 |
| 574 | Height Gage | 6 | 936937 |
| 193 | Digimatic Height Gage | 3 | 905338 |
| 342 | Crimp height micrometer | 2 | 05CZA662 |
| Other | 2D Linear Height Gage | 1 | RS-232, DB9 female, null modem |
Depth gages 55
| Series | Gage types | Models | Connection |
|---|---|---|---|
| 571 | Depth Gage, Point-Type Digimatic Depth Gage | 28 | 959149, 05CZA624 |
| 547 | Depth Gage | 17 | 905338, 06AGL011 |
| 329 | Depth micrometer | 9 | 05CZA662, 937387 |
| 527 | Vernier Depth Gage | 1 | 1 without data output |
Other gages 49
| Series | Gage types | Models | Connection |
|---|---|---|---|
| 192 | Gage | 8 | 8 without data output |
| 547 | Fixture, Bench Gage | 8 | 905338; 6 without data output |
| 187 | Digital Protractor | 4 | 905338 |
| 950 | Digital Protractor | 4 | PRO3600; 2 without data output |
| VL | Digital Measuring Unit | 4 | RS-232, DB9 female, null modem |
| Other | 2-D Data Processing Unit, Multiplexer, Digital Protractor | 3 | RS-232, DB9 female, null modem, RS-232, DB9 female, RS-232, 10-pin T&B Ansley 609-1027 |
| 519 | Stand, Unknown | 3 | 936937; 2 without data output |
| MIG | Gage Multiplexer | 3 | RS-232, DB9 female |
| 176 | Microscope | 2 | 959149 |
| 318 | Litematic | 2 | 936937 |
| 515 | Snap Gage | 2 | 936937 |
| 575 | Linear Gage | 2 | 936937 |
| 518 | RS-232 | 1 | |
| IT | Digimatic to RS-232 Input Tool | 1 | RS-232, DB9 |
| RA | Roundness/Cylindricity Tester | 1 | RS-232, DB9 female, null modem |
| RS | Roundness/Cylindricity Tester | 1 | RS-232, DB9 female, null modem |
Thickness gages 39
| Series | Gage types | Models | Connection |
|---|---|---|---|
| 547 | Thickness Gage | 39 | 905338, 06AGL011 |
Hardness and surface testers 20
| Series | Gage types | Models | Connection |
|---|---|---|---|
| 178 | Surftest | 8 | 936937 |
| SJ | Surftest, Surface Roughness Tester | 8 | 936937, RS-232, 6-pin modular (RJ-12) |
| 811 | Durometer | 4 | 905338 |
Displays and counters 18
| Series | Gage types | Models | Connection |
|---|---|---|---|
| 542 | Counter/displays, Linear Gage Counter | 7 | 936937, RS-232, Digimatic, RS-232, DB9 female, null modem |
| 547 | Chamfer Chek | 4 | 905338 |
| Other | Linear Gage Counter, Display Unit for EV Counter | 2 | RS-232, DB9 female, null modem, RS-232, DB9 male |
| 568 | Display | 2 | |
| 572 | Display | 2 | 936937 |
| 983 | Digital Display Unit | 1 | 1 without data output |
Scales and balances 5
| Series | Gage types | Models | Connection |
|---|---|---|---|
| 174 | Display Unit for Linear Scale | 2 | RS-232, DB25 female |
| 579 | Linear Scale | 1 | 1 without data output |
| KA | Display Unit for Linear Scale | 1 | RS-232, DB9 male |
| Other | Display Unit for Linear Scale | 1 | RS-232, DB25 female |
Force and torque gages 1
| Series | Gage types | Models | Connection |
|---|---|---|---|
| 983 | Torque Wrench | 1 |
Wireless or wired
Wireless (MobileCollect) suits gages that move between machines or travel with the operator. A Mini Mobile Module EVO with the matching connector mounts on the gage and sends each reading to a base receiver at the computer, up to 140 feet away, on one CR2032 battery for 800,000 or more readings. For Mitutoyo micrometers with the 5-pin Digimatic port, the Shape Mic 01 mounts directly on the port with no cable at all, and the Shape Mic 02 does the same for the newer QuantuMike.
Wired (GageWay) suits fixed stations. The Mitutoyo SPC cable plugs straight into a GageWay interface, with no batteries and no pairing. For one gage, the GageWay KW types each reading into Excel or any open program like a keyboard, and the GageWay SM sends it to SPC software over a USB serial port. For several gages at one computer, the GageWay Pro takes 2, 4 or 8 gages, and RS-232 instruments through the GW-LC level converter. The Digital Remote is a middle option: it takes the same Mitutoyo cable at a stand or fixture, is USB powered, and sends the reading wirelessly.
Both systems can run side by side. If you already own Mitutoyo U-WAVE-T, U-WAVE-TM or U-WAVE-TC transmitters (not the Bluetooth models), a MobileCollect base accepts their readings too, so they keep working next to MobileCollect transmitters on other gages and other brands. Keep in mind that U-WAVE was built as a simpler system: Mitutoyo rates its range at about 65 feet, compared with about 140 feet for MobileCollect, and U-WAVE transmitters do not offer MobileCollect settings such as continuous and TIR read modes or remote triggering. For new gages, a MobileCollect transmitter gets the longer range and the full feature set. See U-WAVE compatibility.
Recent Mitutoyo changes to know about
- New connectors on ID-C, ID-F and QuantuMike. NX ID-C and ID-F indicators take the 06AGL011 cable instead of the 905338, and QuantuMike models ending in -40 take the 06AGL111 instead of the 05CZA662. Details in Mitutoyo changed the SPC cable on ID-C, ID-F and QuantuMike gages.
- Digimatic S1. S1 is two-way: with an EVO base and GageSync, your software can read the gage’s serial number and calibration dates. See GageSync and TriggerSync.
- Windows and the USB Input Tool. A 2026 Windows update blocked the driver behind Mitutoyo’s IT-016U in virtual COM mode. See what to do if your USB Input Tool stopped working.
- DP-1VA printers. MobileCollect can feed readings from several gages, including non-Mitutoyo gages, into a Mitutoyo DP-1VA. See wireless multi-gage input for the DP-1VA.
Questions
No. Digimatic is a synchronous signal, not RS-232, so a serial adapter cannot decode it. It needs an interface that reads Digimatic, such as a GageWay, a Digital Remote, a MobileCollect transmitter or Mitutoyo’s own USB Input Tool. How to connect a digital gage to your computer compares the options and costs. Search your model number in the System Builder. It lists the cable or connector for that exact model, or tells you the model has no data output. The gage-side connector depends on the port, so match the cable to the model, not the gage type. No. One MobileCollect base receives readings from every transmitter, whatever the gage, and a GageWay Pro takes calipers, micrometers and indicators on the same unit. Only the gage cable or connector changes from gage to gage. Yes. Trigger the reading with the read button on the transmitter, a foot or hand switch on a Digital Remote or GageWay Pro, or a command from your software. MobileCollect transmitters can also send continuous or TIR (total indicator reading) values. Yes. MicroRidge is a certified Mitutoyo distributor and sells the Mitutoyo SPC cables used with GageWay interfaces and the Digital Remote, along with the MobileCollect connectors for each Mitutoyo port.Can I plug a Mitutoyo SPC cable into a USB-to-serial adapter?
How do I know which Mitutoyo cable I need?
Do I need a different system for each Mitutoyo gage type?
Can I collect data from a Mitutoyo indicator that has no data button?
Is MicroRidge a Mitutoyo distributor?
Every Mitutoyo model in the builder
Select a model to open it in the System Builder with the transmitter, cable and base for your computer. Calipers, micrometers and indicators are listed on their own pages.
Bore gages
339 series: 339-101, 339-102, 339-291, 339-292, 339-293, 339-294, 339-301, 339-302, 339-303, 339-304
345 series: 345-250-10, 345-250-30, 345-251-10, 345-251-30, 345-350-10, 345-350-30, 345-351-10, 345-351-30
468 series: 468-131, 468-132, 468-133, 468-134, 468-135, 468-136, 468-137, 468-138, 468-139, 468-140, 468-141, 468-142, 468-143, 468-144, 468-145, 468-146, 468-147, 468-148, 468-149, 468-150, 468-151, 468-161, 468-162, 468-163, 468-164, 468-165, 468-166, 468-167, 468-168, 468-169, 468-170, 468-171, 468-172, 468-173, 468-174, 468-175, 468-176, 468-177, 468-178, 468-179, 468-180, 468-181, 468-231, 468-232, 468-233, 468-234, 468-235, 468-236, 468-237, 468-238, 468-239, 468-240, 468-241, 468-242, 468-243, 468-244, 468-245, 468-246, 468-247, 468-248, 468-249, 468-250, 468-251, 468-261, 468-262, 468-263, 468-264, 468-265, 468-266, 468-267, 468-268, 468-269, 468-270, 468-271, 468-272, 468-273, 468-274, 468-275, 468-276, 468-277, 468-278, 468-279, 468-280, 468-281, 468-956, 468-957, 468-961, 468-962, 468-963, 468-964, 468-965, 468-966, 468-967, 468-968, 468-969, 468-970, 468-971, 468-972, 468-973, 468-974, 468-975, 468-976, 468-977, 468-978, 468-979, 468-980, 468-981, 468-982, 468-983, 468-984, 468-985, 468-986, 468-987, 468-988, 468-989, 468-990
526 series: 526-007M
543 series: 543-310B, 543-311B, 543-312B
568 series: 568-294, 568-295, 568-296, 568-297, 568-331-10, 568-332, 568-333, 568-334-10, 568-335-10, 568-336-10, 568-337-10, 568-338-10, 568-339-10, 568-340-10, 568-341-10, 568-342-10, 568-343-10, 568-344-10, 568-345-10, 568-346-10, 568-347-10, 568-348-10, 568-361, 568-362, 568-363, 568-364, 568-365, 568-366, 568-367, 568-368, 568-369, 568-370, 568-371, 568-372, 568-373, 568-374, 568-375, 568-431, 568-432, 568-433-10, 568-434-10, 568-435-10, 568-436-10, 568-437-10, 568-438-10, 568-439-10, 568-440-10, 568-441-10, 568-442-10, 568-443-10, 568-444-10, 568-445-10, 568-446-10, 568-447-10, 568-448-10, 568-461, 568-462, 568-463, 568-464, 568-465, 568-466, 568-467, 568-468, 568-469, 568-470, 568-471, 568-472, 568-473, 568-474, 568-475, 568-924, 568-925, 568-926, 568-927, 568-928, 568-929, 568-930, 568-936, 568-955, 568-956, 568-957, 568-958, 568-959, 568-965, 568-966, 568-967, 568-968, 568-969, 568-971-10, 568-972-10, 568-973-10, 568-974-10, 568-975-10, 568-976-10, 568-977-10, 568-978-10, 568-979-10, 568-980-10, 568-981, 568-982-10, 568-983-10, 568-984-10, 568-985-10, 568-986-10, 568-987-10, 568-988-10, 568-989-10, 568-990-10, 568-991-10, 568-992-10
Height gages
Other: LH 600E
192 series: 192-613, 192-614, 192-615, 192-630, 192-631, 192-632, 192-633, 192-655, 192-656, 192-657, 192-658, 192-663, 192-664, 192-665, 192-666, 192-667, 192-668-10, 192-669, 192-670, 192-671, 192-672, 192-673
193 series: 193-630-10, 193-631-10, 193-632-10
342 series: 342-271-30, 342-371-30
515 series: 515-354, 515-355, 515-356, 515-357, 515-358, 515-359, 515-374, 515-375, 515-376, 515-377, 515-378, 515-379
518 series: 518-221, 518-223, 518-230, 518-231, 518-232, 518-233, 518-234, 518-235, 518-237, 518-240, 518-241, 518-242, 518-243, 518-244, 518-245, 518-246, 518-247, 518-321-20, 518-321-20A, 518-341A-21, 518-342A-21, 518-351-A21, 518-352-A21, 518-352A-21
570 series: 570-202, 570-203, 570-205, 570-206, 570-212, 570-213, 570-214, 570-215, 570-216, 570-223, 570-225, 570-226, 570-227, 570-228, 570-229, 570-230, 570-233, 570-234, 570-235, 570-236, 570-240, 570-241, 570-242, 570-243, 570-244, 570-245, 570-246, 570-247, 570-248, 570-302, 570-304, 570-312, 570-313, 570-314, 570-322, 570-324, 570-402, 570-404, 570-412, 570-413, 570-414
574 series: 574-110-1A, 574-111-1A, 574-112-1A, 574-210-1A, 574-211-1A, 574-212-1A
Depth gages
329 series: 329-250-10, 329-250-30, 329-251-10, 329-251-30, 329-350-10, 329-350-30, 329-351-10, 329-351-30, 329-711
547 series: 547-211, 547-211A, 547-212, 547-212A, 547-217, 547-217A, 547-217S, 547-218A, 547-218S, 547-251, 547-251A, 547-252, 547-252A, 547-257A, 547-257S, 547-258A, 547-258S
571 series: 571-100MOT-10, 571-200MOT-10, 571-201-20, 571-202-20, 571-203-20, 571-204-10, 571-205-10, 571-206-10, 571-207-10, 571-211-20, 571-212-20, 571-213-10, 571-214-10, 571-215-10, 571-216-10, 571-217-10, 571-251-10, 571-252-10, 571-253-10, 571-261-10, 571-262-10, 571-263-10, 571-264-10, 571-265-10, 571-301-10, 571-302-10, 571-311-10, 571-312-10
Thickness gages
547 series: 547-300, 547-300A, 547-300S, 547-301, 547-301A, 547-312A, 547-312S, 547-313, 547-313A, 547-315, 547-315A, 547-316, 547-316A, 547-316S, 547-320A, 547-320S, 547-321, 547-321A, 547-360, 547-360A, 547-361A, 547-361S, 547-400A, 547-400S, 547-401, 547-401A, 547-500, 547-500S, 547-512, 547-512A, 547-516, 547-516A, 547-520, 547-520S, 547-526, 547-526S, 547-528, 547-561, 547-561S
Other gages
Other: 264 Series, older MIG-4/MIG-8, Pro 3600
176 series: 176-820A, 176-821A
187 series: 187-501, 187-502, 187-551, 187-552
318 series: 318-211A, 318-213A
518 series: 518-313A
519 series: 519-621
IT series: IT-007R
MIG series: MIG-2USB, MIG-4USB, MIG-8USB
RA series: RA-220
RS series: RS-232-120
VL series: VL-50A, VL-50AS, VL-50B, VL-50BS
Hardness and surface testers
178 series: 178-561-01A, 178-561-02A, 178-563-02A, 178-565-02A, 178-581-01A, 178-581-02A, 178-583-01A, 178-583-02A
811 series: 811-332, 811-334, 811-336, 811-338
SJ series: SJ-201, SJ-201P, SJ-210, SJ-301, SJ-310, SJ-410, SJ-411, SJ-412
Displays and counters
Other: 542 series, D-EV
542 series: 542-007A, 542-063 / 542-064 / 542-067, 542-071, 542-072, 542-073, 542-074, 542-075
547 series: 547-480, 547-481, 547-482, 547-483
Scales and balances
Other: K Series Counter
174 series: 174-183A, 174-185A
KA series: KA-200 Counter
Force and torque gages
983 series: 983-XXX
Other gages without data output: 192-130, 192-131, 192-132, 192-133, 192-150, 192-151, 192-152, 192-153, 519-109E, 519-109M, 547-033S, 547-034S, 547-035S, 547-053, 547-054, 547-055, 950-315, 950-317. These models have no data output, so no cable or transmitter can read them.
Depth gages without data output: 527-xxx. These models have no data output, so no cable or transmitter can read them.
Scales and balances without data output: 579-XXX. These models have no data output, so no cable or transmitter can read them.
Displays and counters without data output: 983-352. These models have no data output, so no cable or transmitter can read them.